- FoI Number
- 2025-007
- Subject
- Lerwick Dental Practice Report
- Date Received
- 02/04/2025
- Request and Response
-
- Under the Freedom of Information Act, we formally request a copy of the report that Mr. Visocchi submitted to you. (FOI Request sent to Dr. S. Laidlaw)
The report you are requesting was not a separate document, rather part of an email. Please see relevant extract below.
‘This is in relation to the operation status and daily verifications of the two existing autoclave/sterilisers at Lerwick Dental Practice
Steriliser 1 – Ultraclean L2DH-2A-1833
Last on-site service, re-validation, testing – 05/03/24
I am awaiting the engineers’ advice to confirm if the manufacturer’s advice is annual or quarterly maintenance, testing and re-validation is appropriate for this model (they would appear to be tested annually to date)
Validation Evidence Provided;
Fully supplied information from data logger from 01 April 2024 – 16 October 2024. No Failures.
Steriliser 2 - Eschmann Little Sister SES 225B
Last on-site service, re-validation, testing – 04/03/24
I am awaiting the engineers’ advice to confirm if the manufacturer’s advice is annual or quarterly maintenance, testing and re-validation is appropriate for this model (they would appear to be tested annually to date)
Validation Evidence Provided;
Information provided from printer; 10 May – 29 July. One Failure
Some dates only had one cycle which is surprisingly low for a four surgery/5 dentist practice (even with two sterilisers)
Additional verification for these dates only; 20/09/ 24, 23/09/24, 01/10/24
No verification from March or April.
As discussed at the PAG, there is no way of determining which instruments went through which machines
Reminder of Dates
Practice Inspection - 16 October
Mitigations agreed and observed - 07 November’